68
1. Connect to DUT
According to the Figure 8 or 9 and the type of reference sensor, to measure the TC or RTD. Then according to
Figure 6 or 7 to measure the voltage or circuit based on the transmitter’s output type.
2. Create the task
Refer to paragraph 6.2.2.
3. Task setting
Refer to the task setting in paragraph 6.1.1.1.
4. Start the task
For part of the content, refer to Start the task in paragraph 6.1.1.4.
During the task performing, the calibrator simulates and outputs a temperature signal to the DUT, measures and
records the electrical signal output of the DUT. No manually input required.
5. Finish the task
Refer to Finish the task in paragraph 6.1.1.5.
6.2.4 Temperature switch
This type of task is for calibrating temperature switch without sensor. It will use the calibrator’s simulate temperature
output as the sensor.
1. Connect to DUT
Connect the DUT with the measurement and output module of the calibrator, as shown in the Figure 20.