81
Select the graph display button
and the table display button
to switch between the graph display or
table display.
Pressure switch action: When the internal pressure module is used as the standard, it will be automatically run
the test. When the external pressure module is used as the standard manual pressure generation is required and the
switch data will be recorded.
To stop this test, click the back button
.
3. Save the result
Click the save button
to save the test result.
The following items need to be entered: Name and serial number, click OK button
to save the test results.
To exit without saving press the back button
.
To redo the test click the reset button
.
5.1.4 Pressure sensor
Select the “pressure sensor” icon
in the test type list to start the quick test.
1. Connection
Connect the sensor as shown in Figure 5-2.
2. Start the test
Same as Section 5.1.2 Pressure transmitter.