87
performed without manually advancing if the measurement value is the same as the set point value.
★
When enabling
the tapping function, pay attention to the status before/after tapping as prompted by the function screen, and input the
indicated value of the gauge before/after tapping.
4) Repeat Step 2&3 until the entire pressure calibration process is completed;
After the entire calibration process is completed, it will automatically enter the report interface.
5. Task end
The ADT761A provides calibration task data review, sorting and save functions.
After the task is over, the user can choose to see the result of the task through the data view icon or the task data.
The user can also click
to abandon the task data and execute the task again.
Task data can be stored by clicking
in the save interface and the following information can be entered:
Table 5- 3 Task save
Item
Effective value
Description
Operator
Letter, numeral, symbol
Calibration task operator information
Execution time
2000/01/01-2099/12/31
Calibrator task execution date
Ambient temperature
Numeral
Ambient temperature during task
Temperature unit
К, °C, °F
Unit of ambient temperature during task
Ambient humidity
Numeral
Ambient humidity during task
Save as
As found, as left and both
Save pre-adjustment data or post-adjustment data, or both
After saving, return to the task interface.
Summary of Contents for 761A
Page 1: ...761A Automated Pressure Calibrator...
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Page 19: ......
Page 20: ......
Page 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Page 33: ...13 LAN Interface USB port Power supply input Label Battery...
Page 47: ...27 Figure 3 3 Gauge pressure outlet...
Page 151: ...131...