96
When the internal pressure module is selected as the standard, the pressure switch test method is based on
slew rate in the control settings and the rate is gradually attenuated following the change of state for the switch.
When using an external pressure module, the control must be done manually.
Click the stop icon at the bottom right of the screen to end the task.
5. Task end
For the contents, refer to Section 5.2.1 Task end for calibration of the analog pressure gauge.
5.2.6 I/P converter
1. Connection
Connect the calibrator and the I/P converter as shown in Figure 3-21;
Never apply pressure out of the range of the calibrator.
If the loop power supply is enabled (with maximum load of 50mA), check the load capacity.
2. New task
Select the “I/P converter” icon
in the test type list, and input the information of the I/P converter.
Table 5- 7 Task mode I/P converter information
Item
Effective value
Description
Name
Letter, numeral, symbol
Name of the I/P converter
Serial number
Letter, numeral, symbol
Serial number of the I/P converter
Model
Letter, numeral, symbol
Model of the I/P converter