106
3. Task setting
For the contents, refer to Section 5.2.1 Task setting for calibration of the analog pressure gauge.
4. Task start
For some contents, refer to Section 5.2.1 Task start for calibration of the analog pressure gauge.
Set loop power settings as necessary.
5. Task end
For the contents, refer to Section 5.2.1 Task end for calibration of the analog pressure gauge.
5.3 Quick test (for barometric (BP) version)
Click the main menu icon
on the main interface, and then select the icon
on the main menu that pops up to
enter the quick test function, and the calibrator displays the test type list.
5.3.1 Digital barometer
1. Prepare
Select “digital barometer” in the test type list to start the quick test of digital barometer.
2. Connection
When testing the general digital barometer, connect to the calibrator as shown in Figure 5-6.
When testing the VAISALA digital barometer, it needs to connect the calibrator and the communication interface
as shown in Figure 5-7.
3. Start the test
Click the start button
to select the automatic execution to enter the execution parameter setting interface.
Fill in the set point list, the number of cycles, the stroke mode, the number of readings, the reading interval, and
Summary of Contents for 761A
Page 1: ...761A Automated Pressure Calibrator...
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Page 19: ......
Page 20: ......
Page 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Page 33: ...13 LAN Interface USB port Power supply input Label Battery...
Page 47: ...27 Figure 3 3 Gauge pressure outlet...
Page 151: ...131...