120
6. Application
6.1 Unit conversion
Click the menu in the upper right corner on the main interface to select the application, and then enter the
pressure unit converter.
Supports conversion between multiple pressure units.
6.2 Pressure leak test
There are two pressurization ways of pressure leak detection:
a) Calibrator pressurization: Select the internal high or low pressure range, and the calibrator will automatically run
the test.
b) External pressurization: Select an external pressure module and use external pressure generation to run the
leak test.
1. Setup
Select the application in the function menu to enter the pressure leak test.
Click the upper area on the interface to select the range: Select the internal module range or the external pressure
module.
1) If the external pressure module is selected, there is no need to set the pressure target value;.
Select the pressure unit: If the selected unit is not supported by the selected module, it will automatically revert
to the last setting.
Set point
1) There is no need to set this parameter when selecting the external pressure module.