127
8.5 Replacement of the filter at the air pressure output interface and the O-ring
Figure 8- 2 Disassembly of air pressure/high-precision/barometric (BP) version pressure connector
①
Unscrew the connector
②
Push out the stainless steel filter
element with a rod with diameter of
less than 2.5mm
③
After cleaning the filter element, press
it back into the connector
④
Replace the sealing ring