5
Accuracy(1)
0.05%FS
0.02%FS
0.02%FS
(2)
0.02%FS
(2)
0.01%FS
0.01%FS
Stability
0.005%FS or
0.05 Pa
0.005%FS
0.005%FS
0.005%FS
0.003%FS
0.02 hPa
Pressure Type
Differential,
Gauge
Differential,
Gauge
Gauge, Absolute Gauge, Absolute Gauge, Absolute
Absolute
Over Range
Indication
120%
Resolution
6 digits
Measurement Units
Pa, hPa, kPa, mPa, bar, mbar, psi, mmHg@0°C, cmHg@0°C, mHg@0°C, inHg@0°C, inH2O@4°C,
mmH2O@4°C, cmH2O@4°C, mH2O@4°C, mmH2O@20°C, cmH2O@20°C, mH2O@20°C, inH2O@20°C,
kg/m2, mtorr, torr, lb/ft2, tsi, custom
Barometric Accuracy
N/A
N/A
±60 Pa
±60 Pa
±60 Pa
N/A
[1] One year accuracy (including 1 year stability). FS specification applies to the span of the module range.
[2] Specification based on gauge measurement. An additional 60 pa uncertainty will need to be included when measuring