35
Extremum pair number
Integer 0~10
Extremum pair number of the average filter
Resolution: Allows setting digit resolution.
Stability: Allows setting the stability of measurement data.
Loop power supply: Applicable only for the current measurement. Select enabling/disabling loop power supply.
3.3.2 Switch test
1. Similar to 3.3.1, open the electrical measurement options, click the range area on the screen, and select
for switch
test. The calibrator can test three types of switches, i.e. mechanical switch, NPN electronic switch and PNP electronic
switch.
2. Connection
If mechanical switch is selected, connect the electric circuit as shown in Figure 3-7.
If NPN electronic switch is selected, connect the electric circuit as shown in Figure 3-7.
If PNP electronic switch is selected, connect the electric circuit as shown in Figure 3-7
Figure 3- 7 Switch testing
The switch action values are recorded only when the output item is pressure.