45
The device description file is used to describe device parameters and parameter access modes. Through the parameter
description information, you can view and set related parameters of the PA transmitter. The calibrator uses specific device
description files, to access main parameters in the Physical Block, Transducer Block and Function Block of PA transmitter.
This calibrator includes device description files of common mainstream PA pressure transmitters. If you need to add new
device description files of transmitter, please contact us.
2. Connection and search
For the connection mode, please refer to Figure 3-10.
Click the
icon to start searching; and click the
icon to stop searching.
From the searched PA device list, click a PA device to be connected. After connection is successful, return to the
main interface.
3. Process quantities
Click
, then a function menu will pop up. Select "Process Quantity" in the function menu. The calibrator provides
switching between display of process quantities such as PRIMARY_VALUE, SENSOR_VALUE, SECONDARY_VALUE_1,
TRIMMED_VALUE, SECONDARY_VALUE_2 and STATIC_PRESSURE_VALUE.
4. Transmitter operation
(1) Setup
After clicking
, a function menu will pop up. Select "Setup" in the function menu to enter the setup interface.
On the setup interface, you can access and set parameters in the Physical Block, Transducer Block and Function
Block.
Before modification of some parameters, you may need to modify corresponding TARGET_MODE (e.g. set it to
Summary of Contents for 761A
Page 1: ...761A Automated Pressure Calibrator...
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ...X Table 5 13 Task mercury barometer 120 Table 5 14 Task barometric altimeter 123...
Page 19: ......
Page 20: ......
Page 31: ...11 2 Installation 2 1 Features 2 1 1 Basic structure Figure 2 1 Basic structure...
Page 33: ...13 LAN Interface USB port Power supply input Label Battery...
Page 47: ...27 Figure 3 3 Gauge pressure outlet...
Page 151: ...131...