86
5.2.1 Dial pressure gauge
1. Connection
Connect the calibrator and the dial pressure gauge as shown in Figure 20.
2. New task
Select the “Dial pressure gauge” icon
in the test type list and input the information of the pressure gauge.
Table 25 Task mode dial pressure gauge information
Subject
Valid value
Description
Name
Letter, numeral, symbol
Name of the pressure gauge
Serial number
Letter, numeral, symbol
Serial number of the pressure gauge
Model
Letter, numeral, symbol
Model of the pressure gauge
Pressure type
Gauge pressure, absolute pressure,
differential pressure
Pressure type of the pressure gauge
Range
Depends on the pressure gauge
Range and unit of the pressure gauge
Accuracy
0.06%, 0.1%, 0.16%, 0.25%, 0.4%,
0.6%, 1%, 1.6%, 2.5%, 4%, and
custom
Precision of the pressure gauge. The numeric input in the custom option
is the accuracy of the pressure gauge. For example: For 1.5% accuracy
would be input as 1.5 through the numeric keyboard, which has a range
of (0.001 ~ 100)%
Resolution
Minimum resolution: 0.001
The resolution of the pressure gauge
Owner
Letter, numeral, symbol
Owner of the pressure gauge