87
3. Task setting features
Select the task in the task list to enter the instrument parameter interface, click the start icon to enter the execution
information interface, and input the calibration information of the pressure gauge.
Table 26 Task setting of dial pressure gauge
Location
Letter, numeral, symbol
Location position of the pressure gauge
Note
Letter, numeral, symbol
Notes of the pressure gauge
Subject
Valid value
Description
Set point
s
The Valid value of the set point
pressure depends on the pressure
range of the gauge; up to 17
calibration points can be set
Set the calibration points for the task, and the calibrator will
automatically set the default calibration point based on the range of
the calibrated instrument and the number of the calibration points. Any
point can be manually selected to be changed to a different point. The
set point needs to be within the range of the pressure gauge. Change
the number of the calibration points through the arrow key on the right
side of the screen or by clicking the number of the calibration points.
Cycle number
1, 2, 3
Select the number of cycles
Stroke
Round trip / one way
Select the calibration task stroke mode
Dwell time
Integer of (1 ~ 600) seconds
The waiting time to take the readings before the next set point.
Tapping
Enable, disable
When enabled the value of the dial pressure gauge needs to be input
twice for each calibration point, and then the calibrator will shift to the
next calibration point to continue the task.