117
Figure 34 Reservoir Assembly Maintenance
Rotate the two hydraulic nuts counterclockwise respectively, remove them, check the sealing ring on the
hydraulic interface, and replace it if it is damaged, as shown at position
④
;
Rotate the vent valve nut counterclockwise to remove it, as shown at position
⑤
;
Take out two liquid line filters and liquid injection filters, respectively, for cleaning or replacement, as shown in
position
⑥
in the figure. The liquid line filters and liquid injection filters have different sizes and cannot be
mixed;
Clean or replace the liquid storage tank and replace it with new media;
3. Reservoir Assembly Installation
Reverse the steps in 1 and 2 to assemble the reservoir assembly.
Note: When installing the hydraulic interface, the heights of the two hydraulic interfaces can be adjusted appropriately to
suit the height of the corresponding joint of the whole machine.