The set value shall meet the calibrator's current output range of 0~25mA, voltage output range of 0~16V, and
power supply output range of 16~30V.
5. Auto step
Press the
key to enter the auto step setup. For the auto step parameters, see Table 14.
Table 14 auto step parameters
Subject
Valid value
Description
Stroke
Round trip or one way
Set the trip pattern of auto step.
Cycle number
0-100
Set the cycle number of auto step.
Step mode
Step point, percentage, value,
or custom
Set the mode of automation step.
Cycle interval
0~3600 sec
Standing time between the end of each cycle and the start of next cycle.
Dwell time
0~3600 sec
Standing time after the present electrical output becomes stable.
Step point number
2-17
It can be set when the step mode is step point.
Range
Not to exceed the electrical
output range.
Set the output range of automatic step.
Step point list
Editable when the step mode
is custom; read-only and
displayed under other modes.
Display the set point list of automatic steps.
Step value
6.25%~100%
It can be set when the step mode is percentage.
Step value
1.5625~25mA
It can be set when the step mode is value.