57
Figure 25 Calibration of a pressure switch
3. Capture of the action value
To capture more accurate action values, you can enter "control settings” and set the slew rate to a lower value
which will reduce the lag time.
Separately use the upper and lower range limits of switch as the target value for pressure control until the switch
acts. Then, capture and display the action value.
4. Record the captured action value or use snapshot storage (it can store a pair of action values at the same time).
3.6.5 I/P converter
1. Prepare