71
pressure is stabilized, click
to do the next point calibration.
Note: Please adjust the reference gauge pressure output to be the same as the set value of the calibrator.
If the measured value deviates too much from the calibration point, there will be a prompt
After the calibration is successful, the new calibration data will take effect immediately, and the calibration date
will become the current system date
4.
Restoring factory calibration data
Click on the right side of the screen the
button can be restored to factory calibration data.
4.4.5 Calibration Expiration Reminder
You can set the calibration expiration date of each module and set whether to turn on the expiration reminder function of
each module. The default calibration period is 365 days. When a specific module of the calibrator has exceeded the
calibration period, a prompt will be given when the calibrator is powered on or when the module is online . You can turn
on/off the reminder function or modify the calibration period in the calibration expiration reminder settings interface
4.5 Services
4.5.1 Firmware upgrade
The calibrator provides a function of firmware upgrade. The upgrade operation includes two types: Local upgrade via a
USB flash drive or remote upgrade.
1.Local upgrade via a USB flash drive
Summary of Contents for 762