47
1111
Pressure module: constant voltage excitation failed
1121
Pressure module: internal EEPROM failed
1131
Pressure module: external EEPROM failed
Note: In case of it cannot be turned on, please firstly open the cover of the pressure gauge, replace the battery and press the power
button to turn it on again. If it still, please contact the manufacturer.
3.9 Operation Error Codes
When performing function operations or information editing operations on the screen, since the screen cannot display
detailed text, the pressure gauge will prompt in the form of an error codes.
Table11 Operation Error codes list
Error codes
Description
Measure
001
Password wrong
Check the password
002
Parameter beyond the range
The entered value is beyond the
gauge’s range, enter a new value
003
Operation execution failed
Try it again
004
Flash initialization failed when performing data
recording
Restore to factory settings, if the fault
still, please contact the manufacturer