7
90612-30-180320-Rev009-UM-en
By pressing
you’ll pass to next unit. Please repeat this step for each unit.
After setting the last unit (oz) and finalising with
, display will show CAL (with G value).
In this parameter you must not make any changes, because otherwise it’ll change metrological
parameter which could lead to wrong measuring results.
Just press
.
The scale turns automatically in weighing mode and is now ready for weighing with new
parameter settings.
After switch-off the scale parameter settings will be stored.
Cleaning & Maintenance
Cleaning
Before cleaning make sure the scale has been disconnected from the socket. For cleaning just
use a damp cloth or common disinfection products. Do not use any aggressive cleaning fluids.
Make sure no water or cleaning fluid penetrates the scale, because it could
damage the scale irreparably!
The scale must not be cleaned with high-pressure cleaner!
Display- and Error indication
Symbol
Indication
Reload accumulator with IEC Power Connector.
--OL--
Overload: Scale ist overloaded, the weight of object is higher than capacity.
Remove the object immediately from the weighing platform.
------
Underload: If there’s a container on the platform when switch-on the scale and if
then the container will be removed, from the platform, display will show an
underload and an acoustic warning signal appears. Switch-off the scale and
switch-on again for following up weighing process.
Err 4
Zero error: The scale is above zero range (4% of maximum). Remove the object
from the weighing platform and restart the scale. If the error still occurs, contact
your local service partner or dealer for re-calibrating the scale.
Err 5
Keyboard error.
Contact your local service partner or dealer.
Err 6
A/D-converter out of
regularly range
Safety instructions
Do NOT use the scale
:
If the mains plug or cable shows visible damages
If the IEC Power Connector is not working regularly.
After long storage in humid places