Seite
36
von
112
M118600-01-210305-Rev002-UM-en
Magnetic field at the supply
frequency (50/60 Hz)
according to IEC 61000-4-8
30 A/m
30 A/m
The line-frequency magnetic fields
should correspond to the
characteristics of a typical installation
site in a commercial or clinical
environment.
NOTE:
UT is the alternating mains voltage prior to the application of the testing level.
Guidance and manufacturer’s declaration - electromagnetic immunity
This PRODUCT is intended for use in the electromagnetic environment specified below. The customer or the user of
the PRODUCT should make sure that it is used in such an environment.
Immunity Tests
Immunity
Tests
Immunity
Tests
Immunity Tests
HF conducted
disturbances as
per
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Portable and mobile RF communications equipment should be used
no closer to any part of the PRODUCT, including cables, than the
recommended separation distance calculated from the equation
applicable to the transmitter frequency.
Recommended protective distance:
= 1.2√
HF radiated
disturbances as
per
IEC 61000-4-3
10 V/m
80 MHz to
2.7 GHz
10 V/m
=
6
√
= 1.2√
80 Mhz to 800 Mhz
= 2.3√
800 MHz to 2.7 GHz
Where (P) is the maximum output power of the transmitter in watts (W)
according to the transmitter manufacturer's specifications and d is the
recommended separation distance in metres (m).
The field strength of fixed RF transmitters, as determined by an
electromagnetic site survey a should be less than the compliance level in
each frequency range.b
In the vicinity of devices that bear the following pictorial symbol,
interference is possible:
COMMENT 1
At 80 MHz and 800 MHz, the higher value shall apply.
COMMENT 2
This guidance may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
The field strength of fixed transmitters such as base stations for wireless telephones and mobile land radio
services, amateur radio stations, AM and FM radio and television transmitters cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed HF transmitters, an electromagnetic site survey
should be considered. If the field strength measured at the location of the PRODUCT exceeds the compliance level
indicated above, the PRODUCT should be monitored for its normal operation at each place of use. If abnormal
performance characteristics are observed, additional measures may be necessary such as reorienting or relocating
the PRODUCT.
b
In excess of the frequency range 150 kHz to 80 MHz, the field strength should be less than 3 V/m.