M320600-210429-Rev004-UM-en
Voltage drops, short-
term interruptions, and
fluctuations of the
supply voltage
according to
IEC 61000-4-11
0% UT for 1/2 period
(100% break-in)
0% UT for 1 period
(100% break-in)
40% UT for 5 periods
(60% break-in)
70% UT for 25 periods
(30% break-in)
80% UT for 250 periods
(20% break-in)
0% UT for 250 periods
(short interruption)
0% UT for 1/2 period
(100% break-in)
0% UT for 1 period
(100% break-in)
40% UT for 5 periods
(60% break-in)
70% UT for 25 periods
(30% break-in)
80% UT for 250 periods
(20% break-in)
0% UT for 250 periods
(short interruption)
The quality of the supply voltage
should correspond to that of a typical
business or hospital environment.
If the user of the PRODUCT requires
continued operation even during power
interruptions, it is recommended that
the PRODUCT be supplied by an
uninterruptible power source or a
battery.
Magnetic field at the
supply frequency
(50/60 Hz) according
to IEC 61000-4-8
30 A/m
30 A/m
The line-frequency magnetic fields
should correspond to the
characteristics of a typical installation
site in a commercial or clinical
environment.
NOTE:
UT is the alternating mains voltage prior to the application of the testing level.
Guidance and manufacturer’s declaration - electromagnetic immunity
This PRODUCT is intended for use in the electromagnetic environment specified below. The customer or the user of
the PRODUCT should make sure that it is used in such an environment.
Immunity Tests
IEC 60601 Test Level
Compliance
level
Electromagnetic environment – guidelines
HF conducted
disturbances as
per IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Portable and mobile RF communications equipment
should be used no closer to any part of the
PRODUCT, including cables, than the recommended
separation distance calculated from the equation
applicable to the transmitter frequency.
Recommended protective distance:
= 1.2√
HF radiated
disturbances as
per IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
=
6
√
= 1.2√
80 Mhz to 800 Mhz
= 2.3√
800 MHz to 2.7 GHz
Where (P) is the maximum output power of the
transmitter in watts (W) according to the transmitter
manufacturer's specifications and d is the
recommended separation distance in metres (m).
The field strength of fixed RF transmitters, as determined
by an electromagnetic site survey a should be less than
the compliance level in each frequency range.b
In the vicinity of devices that bear the following
pictorial symbol, interference is possible: