M403660-211129-Rev006-UM-e 

 

Print out a record 

The scales must be at standstill. The standstill status symbol 



 must appear at the screen. Press the 

ENTER key to send the data to the serial interface.  

Note: Printing is not possible if the net weight is 0. The print function only produces one print-out. A reset 
is required to print out another copy.  

Error Messages 

Error Codes 

Description 

Solution 

 

 

Load cell wires to indicator are 
incorrectly connected; or shorted, or 
opened; or ADC/ load cell are being 
damaged  

Make sure wires are in good order 
and condition and correctly 
connected. Replace load cell or 
ADC chip. Service is required.

 

 

 

Weight reading exceeds the zero limit 
after switching the scale on. 

Make sure scale platform is empty. 

 

Weight reading is below the zero limit 
after switching the scale on. 

Install platform on scale. 
Otherwise the zero calibration/ 
adjustment have to be done by 
authorized personnel. 

 

Weight reading exceeds load limit.  

Reduce load on scale until weight 
value can be displayed.  

 

Weight reading below load limit. 

Service by authorized personnel 
required. 

 

Configuration or Calibration parameters 
are not correctly set. 

Contact your local dealer / supplier 
for calibration purpose. 

 

Capacity parameters are not correct. 

Contact your local dealer / supplier 
for a new set-up. 

 

Calibration error. The incoming signal of 
loaded weight is too small, too big, 
unstable or un-linear. 

Contact your local dealer / supplier 
for a new set-up. 

Not turn on. 

Power cord is not properly connected. 
Power outlet not supplying electricity. 
Batterie is discharged. Other failure. 

Check power cord connections. 
Make sure power cord is plugged 
into the power outlet. Check power 
source. Replace batteries. 

Cannot zero the 

display or will 

not zero when 

turned on. 

Load on scale exceeds allowable limits. 

Load on scale is not stable. 

Load cell damage. 

Remove load from the scale. 

Wait for load to become stable. 

Service required. 

No Battery 

symbol or 

display of Lo.bAt  

Batteries are discharged. 

Charge batteries. 

 

Summary of Contents for M403660

Page 1: ...Operating Manual International Range Electronic Chair Scales M403660 M403660 211129 Rev006 UM e...

Page 2: ...scription 4 Key pad 5 Operation 5 Positioning of patient 5 How to change the unit 6 Tare function 6 Weighing with preset tare values 6 HOLD function 7 BMI Body Mass Index 7 Evaluation of the Body Mass...

Page 3: ...uctions in this operating manual the country specific installation standards as well as the applicable safety conditions and accident prevention regulations in all aspects Operate the scale only with...

Page 4: ...rgeable batteries Use the AC adapter to load the rechargeable batteries The imprinted voltage value at the mains adapter must match the local mains voltage When the battery is being charged for the fi...

Page 5: ...thin the zero level Standstill status indicator 14 The scale is at a standstill meaning the weight is stable and no movement is on the platform This is a mandatory precondition for functions such as r...

Page 6: ...witch between the numbers downwards In BMI mode use this key to switch between the numbers upwards Enter button to confirm the value Operation Positioning of patient Please ensure for security reasons...

Page 7: ...hing on to get proper result Tare function The tare function allows ignoring a fixed weight e g clothes that will be deducted after each weighing procedure Proceed as follows Press the start key with...

Page 8: ...w buttons to enter the height of the patient Confirm the height with the enter button The BMI is now being displayed To correct the given height figures press the BMI button once again and the display...

Page 9: ...thorized personnel Weight reading exceeds load limit Reduce load on scale until weight value can be displayed Weight reading below load limit Service by authorized personnel required Configuration or...

Page 10: ...elivery applying to defects occurred due to poor material or workmanship presentation of invoice is necessary All moveable parts batteries cables mains units rechargeable batteries etc are excluded Wa...

Page 11: ...dapter AC Adapter 6 9VDC 500mA Central negative 4xAA NIMH rechargeable batteries pack Display 6 digit 7 segment 1 25mm LCD with 16 annunciators and blue backlight Working temperature 10 C to 40 C Stor...

Page 12: ...ermany 49 40 432 776 0 49 40 432 776 10 info ade germany de www ade germany de Declaration of conformity by the manufacturer These scales have been manufactured in accordance with the harmonised Europ...

Page 13: ...M403660 211129 Rev006 UM e 12...

Page 14: ...13 M403660 211129 Rev006 UM e...

Page 15: ...M403660 211129 Rev006 UM e 14...

Page 16: ...15 M403660 211129 Rev006 UM e authorised dealer...

Reviews: