Volt/mA Calibrator
Using
3.2.5 Auto Ramp Function
Auto Ramp function is applied to all of output signal.
In SOURCE mA mode Auto ramp function allows you to apply a varying current singal continuously
from the Calibrator to a passive (sourcing) or active (simulate) loop and you can still leave your hands
free to test
the transmitter’s response.
Similary In SOURCE V/SOURCE mV mode,auto ramp function apply a varying voltage singal
continuously from the Calibrator to load circuit.
Press
to enter into “Auto Ramp ” mode and select a ramp type.
The Calibrator applies or controls a continuously repeating current or voltage signal in one of three
ramp types in the span range.
Step 25% stepping ramp with 5s of pause per step.
Slow ramp: Smooth ramp with rate of change 0%-100%-0% within 40s.
Fast ramp: Smooth ramp with rate of change 0%-100%-0% within 15s.
Press any pushbutton (except
pushbutton) or turn off the Calibrator to exit this function.