ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37
Volt/mA Calibrator
Using
3.2.6 The “SpanCheck” Function (0%-100%)
The “SpanCheck” Function apply to all of output signal.
It realize the current/voltage signal from zero to full scale point transition /step output.
In SOURCE V/SOURCE mV mode,proess key, calibrator can execute 0% point of full scale
output,press key again,it can execute 100% point of full scale output, cycle repeatedly.
“SpanCheck” function can be used to check the zero points or span spacing of a transmitter under
SOURCE or SIMULATE mode.
Press to select “SpanCheck”.
Press any pushbutton (except pushbutton) or rotate the knob to exit this function.
The Calibrator has two mA output spans:
4 mA-20 mA (0%-100% ) [default value] ; 0 mA-20 mA (0%-100%) [optional];
To change the output span, press and simultaneously. The setting value for selection will
be saved until next change.
17