ADEL Instrumentation – 28 Rue de Stalingrad – 38300 BOURGOIN JALLIEU 1 – 04 74 93 06 37
Volt/mA Calibrator
Technical Index
4.4 SOURCE / SIMULATE mA dc
Range
Resolution
Accuracy
0 ---- +24 mA
1 uA
±
(
0.02% of 4 Counts
)
SOURCE mode with load ability:Compliance: To 1000
Ω
at 20 mA.
External loop voltage specification under SIMULATE mode: Typically 24V, max 30V, min 12V.
4.5 SOURCE V dc
Range
Resolution
Accuracy
0 ---- +10V
1mV
±
(
0.02% of 4 Counts
)
4.6 SOURCE mV dc
Range
Resolution
Accuracy
0 ---- +100 mV
0.01 mV
±
(
0.02% of 4 Counts
)
4.7 Loop Power:
≥
24V
,≤
26V
.
21