User Manual MKA 500 / MKA 800
Conforms to
Issue 21.02.2023_01 Page 39
8.6 Data logger
MKA 500
MKA 800
During operation the controller can send the current program data to a data logger
(separate
hardware is needed DLK-1-Kit)
.
DLK-1-Kit consists of: Additional board ZSU, µSD card, 3-part USB socket port.
Data Logging Active: Display of the currently pre-programmed status
„No“ = Data logger is inactive
„Yes“ = Data logger is active
„Auto“ =Automatic logging of data: Processes are automatically logged
onto a µSD card, when VisuNet is not active
Interval: display the time intervals at which you want the data to be logged.
Factory setting:
"Auto"= automatic logging of data
Interval = 000 min.
Hint
At a pre-set interval time of 0 min, the data is logged at 30-second intervals.
If the data logger is activated (Active = Yes), there may be a slight delay of up to 3
seconds at the start of a program.
Hint
The data logger can only be used when VisuNet has been enabled and a µSD card
with sufficient memory has been inserted in the relevant recess.
3 Sec.
Profile
Datalogger
3 Sec.
Profile
Datalogger
1
Current time
Enabling data logger
Code number
Interval times in minutes
1
2
2