34
Operation Theory
Positive-slope trigger event occurs
External Digital Trigger
An external digital trigger occurs when a rising edge or a falling
edge is detected on the digital signal connected to pin3 of JP1
for external digital trigger.
Simultaneous Trigger for Multiple Cards
When multiple PCI-9812 cards are used in one system, the trigger
sources of every card can be connected together to provide the
function of simultaneous trigger for multiple cards. Please note
that simultaneous trigger is not equivalent to simultaneous A/D
conversion. The theoretical time difference between the samples
on different card will be ½ of the clock period. Refer to section 5.4
for more information about A/D conversion clock control.
Summary of Contents for NuDAQ PCI-9810