62
Calibration
7.3 A/D Calibration
AD Calibration for Channel 0
1. Apply a +1V input signal to A/D channel 0, and trim the
VR5 until the average reading of channel 0 is within the
range of 2046.6
±
0.1 (9812) or 510.9
±
0.1 (9810).
2. Apply a +0V input signal to A/D channel 0, and trim the
VR1 until the average reading of channel 0 is within the
range of
±
0.2.
3. Repeat steps 1 and 2, adjust VR5 and VR1.
AD Calibration for Channels 1,2,3
The other channels can be calibrated using the steps shown in
section 7.3 by trimming the corresponding VRs. Please refer to the
following table when calibrating other channels.
A calibration utility is provided in the software CD, which is
included in the package. Details of the calibration procedures and
description can be found in this utility. Simply follow the
instructions in the software calibration utility. The measurement
data can also be found within.
Users do not need to calibrate the PCI-9812/10 when using for the
first time as ADLINK has already calibrated it before shipping.
Channel
Number
Full Scale
Adjustment VR
Offset Adjustment VR
0
VR5
VR1
1
VR6
VR2
2
VR7
VR3
3
VR8
VR4
Table 7-2: AD Calibration for Channels 1,2,3