Calibration
61
7
Calibration
In data acquisition, calibrating your measurement devices to
maintain the accuracy is very important. Users can calibrate the
analog input and analog output channels under the users'
operating environment to optimize the accuracy. This chapter
describes the calibration process for PCI-9812/10.
7.1 What you need
The following equipment would be required for the process:
X
Calibration utility: The program would guide the user
through the calibration process once executed. This
program is included in the delivered package.
X
A voltage calibrator or a very stable and noise free DC
voltage generator.
7.2 VR Assignment
There are eight variable resistors (VR) on the PCI-9812/10 board
that allows the user to make accurate adjustments on A/D
channels. The function of each VR is specified in Table 7-1:
VR1
A/D channel 0 offset adjustment
VR2
A/D channel 1 offset adjustment
VR3
A/D channel 2 offset adjustment
VR4
A/D channel 3 offset adjustment
VR5
A/D channel 0 full scale adjustment
VR6
A/D channel 1 full scale adjustment
VR7
A/D channel 2 full scale adjustment
VR8
A/D channel 3 full scale adjustment
Table 7-1: Functions of VRs