background image

UDB97 

Up-Down Ramp Buffer 

Instruction Manual 

1.0  INTRODUCTION 

These instructions refer to the above model.  Supplementary 

sheets are attached if the unit has special options or features. 

For detailed specifications,  see page 4 or refer to the Data 

Bulletin.  All ADTECH  instruments are factory calibrated and 

supplied with a label detailing the calibration. Adjustments are 

normally not necessary. A simple check should be performed 

to verify calibration before installation to ensure that it matches 

the field requirement. 

2.0  GENERAL DESCRIPTION 

The ADTECH  UDB 97 Non-Isolated Up-Down Ramp Buffer 

provides supervisory process control.  A contact change or a 

pulse input ramps the standard control signal output of 1-5 vdc 

or 4-20 ma de up or down or holds the output signal at the 

desired level. 

Two independent contact inputs control the direction of the 

output of the ramp and/or hold function. 

Power  Option  P  1  Non-Isolated  24  vdc  provides  negative 

output signal and negative DC power common rail connection. 

The  output  is a  true  current  source  and  provides  process 

signals  such  as  4-20  ma,  0-1  ma,  0-1 0  ma,  1-5  ma  and 

10-50 ma  de  or  alternatively,  a  voltage  signal  of  5  vdc  full

scale.  Other  current  and voltage 

Inputs/Outputs (1/0) 

are

available as specified on the Data Bulletin.

3.0  INSTALLATION 

The instrument is supplied in a general purpose enclosure as 

standard. 

NEMA 

4, 7 or 12 and plug in chassis enclosures are 

optionally available.  Installation area/location must agree with 

the supplied instruments including operating temperature and 

ambient conditions. 

Mounting 

Refer  to the  appropriate outline  drawing for  mounting  and 

clearance dimensions.  The instrument is surface mounted 

with two #10-32 screws on 8.00 inch centers. 

Electrical Connections 

The wire used to connect the instrument to the control system 

1/0 

should be a twisted pair(s) and sized according to normal 

practice. 

A 12 position barrier terminal block with #6-32 screws and 

3/8"  spacing  is  provided  for 

1/0 

and  power  connection. 

A housing ground terminal marked G is also provided. 

Controls 

Multiturn ZERO, SPAN, UPRATE and DOWNRATE controls 

are provided to calibrate the instrument. The multiturn controls 

are  accessible  through  the  instrument  front  panel  and  are 

clearly marked for ease of use. 

4.0  MAINTENANCE 

These instruments are electronic and require no maintenance 

except periodic cleaning and calibration verification.  If the unit 

appears to be mis-operating it should be checked as installed 

per section 6.0 or removed for a bench  check per sections 

6.0-7.0.  MOST  problems  are traced  to field  wiring  and/or 

associated  circuits.  If the  problem  appears  to  be  with  the 

instrument, proceed to sections 6.0 and 7.0. 

5.0  CONNECTIONS 

Standard connections are shown below and on the instrument 

face plate, Data Bulletin or on attached supplementary sheets. 

RAMP ..r:--

HOLD� 

© 

OUTPUT

(

-) 

© 

(+) 

© 

UP 

__r:--

(j) 

DOWN� 

© 

N  (- FOR DC) 

® 

AC 

(+ 

FOR DC) 

rFP 

-1-

190-A-000029-A

Reviews: