1314-900 Revision BA
250W Precision Calorimeter
27
7
Appendix A
PARAMETER NOMINAL VALUES AND LIMITS
Field Values
Low Limit Value
Reading
High Limit Value
Reading
Units
Expected Values
(250 watts @ 25°C ambient)
THERMISTOR
22
28
°C
26.45°
RTD_TEMP_PRE
20
24
°C
22.75°
RTD_TEMP_POST
20
26
°C
24.45
THERMOPILE
0
0.072
volts
0.066
WATERFLOW_TEMP_LOSS
0
0.002
°C
n/a
SPECIFIC_HEAT
4.18
4.1816
J/g-C
4.18017
GRAVITY
0.9972
0.9982
g/cm3
0.9976
VISCOSITY
0.5
100
mPa-sec
1
FLOWMETER
260
285
pulses/sec
268 to 272
K_FACTOR
8000
8000
pulses/liter
8000
FLOWRATE
0.5
.058
GPM
.51 to .55
FLOW_VOLUME
31.861
34.386
cm3/sec
33.14
MASS_FLOWRATE
31.784
34.302
gm_sec
33.06
SEEBECK_CONSTANT
0.04
0.06
volts/C
0.04
DELTA_T
0
1.8
°C
1.65
LOAD_EFFICIENCY_FACTOR
0
1
--
0.995
FRICTIONAL_WATTS
0
6
watts
4.9
RF_PRESSURE_DROP
15
25
psi
21.094
TOTAL_PRESSURE_DROP
20
30
psi
26.041
INPUT_POWER
-5
250
watts
220 to 260
RF_LOAD_DC_RESISTANCE
45
55
ohms
49.3
SERIAL_NUMBER_OF_RF_LOAD
0
100000
--
0
REFLECTION_COEFFICIENT
0
1
--
0.022
TRANSMISSION_LOSS_FACTOR
0
1
--
0.9992
CAL_FACTOR
-5
250
watts
250
RF_TRANS_LOSS_FACTOR
0
1
--
0.9992
RF_EFFICIENCY_FACTOR
0
1
--
0.9995
USER_DEF_FREQ
0
3500
MHx
0 to 3500
RFCALPOW
-5
250
watts
249.8
POWER_ADJ
-4
4
watts
n/a
T_PIVOT
23
27
°C
25
T_POWER
0.4
0.5
watts/°C
n/a
n/a = not applicable for model #1314