1316-900 Revision AA
10kW Precision Calorimeter
13
Figure 2: Thermocouple Wire Connection
WARNING:
Comply with all local codes and regulations for installation of all
instrumentation. Follow guidelines in supplier manuals for connections to AC Power
Source and 12 kW Chiller. Failure to install equipment properly may result in death or
bodily injury, or equipment damage.
6.
Connect AC mains power to the Calorimeter, AC Power Source, 12 kW Chiller, and
Power Analyzer. The Power Analyzer and Calorimeter are powered from separate 200-
240 VAC outlets. The Calorimeter requires a 20A service.
Chiller Preparation
CAUTION:
Never operate cooling pump dry! Severe damage will occur.
Never apply AC or RF power to the unit until proper coolant levels and sufficient air
flow are established.
CAUTION:
Use only pure distilled water in the coolant mix.
Do not fill the unit with de-ionized water, flammable fluids, corrosive fluids, explosive,
or any other hazardous fluids.