1316-900 Revision AA
10kW Precision Calorimeter
15
9.
On the 12 kW Chiller display, touch the
PUMP ONLY
icon and hold for 2-3 seconds
to circulate coolant.
10.
Allow sufficient time for the coolant to circulate through the system before proceeding
to the next step.
NOTE:
If the coolant level drops and triggers a low coolant alarm, add additional
coolant mixture as necessary to the 12 kW Chiller reservoir until the alarm turns off.
Clear the alarm on the 12 kW Chiller display.
11.
On the 12 kW Chiller display, touch the
PUMP ONLY
icon and hold for 2-3 seconds
to stop flow circulation coolant.
12.
Power OFF the 12 kW Chiller.
Calibration Preparation and Instrument Setup
1.
Install the HPC-CAL calibration software on the system workstation.
2.
Power ON the 12 kW Chiller.
3.
On the 12 kW Chiller display, touch
RUN
and hold for 2-3 seconds.
4.
Power ON the Calorimeter. Once the Calorimeter completes the boot process:
a.
On the Calorimeter front panel, press the
SETUP
softkey.
b.
Navigate to
Instrument
using the arrow keys and press
ENTER.
c.
Navigate to
PUMP
and press
ENTER.
d.
Press
ENTER
again to begin editing the parameter value.
e.
Navigate to
USE PUMP
and press
ENTER.
f.
Press the
UP
arrow key to change the parameter value to
TRUE
and press
ENTER.
5.
Verify that the green
FLOW OK
indicator is lit.
6.
Power ON the 1 kW Chiller.
7.
Connect cables CA-70-5 and CA-71-5 to the Power Analyzer as shown in
Figure 5: CA-70-5 and CA-71-5 connections to Power Analyzer