1316-900 Revision AA
10kW Precision Calorimeter
26
6
Appendix A
PARAMETER NOMINAL VALUES AND APPROXIMATE LIMITS
Field Values
Low Limit Value
Reading
High Limit Value
Reading
Units
Expected Values
(10k watts @ 25°C ambient)
THERMISTOR
22
28
°C
26°
RTD_TEMP_PRE
21.6
22.4
°C
22°
RTD_TEMP_POST
21.6
32
°C
31.5°
THERMOPILE
0
0.8
volts
0.73
WATERFLOW_TEMP_LOSS
0
0.002
°C
0
SPECIFIC_HEAT
3.6
3.8
J/g-C
3.75
GRAVITY
1
1.05
g/cm3
1.03
VISCOSITY
0.5
2
mPa-sec
1.51
FLOWMETER
300
320
pulses/sec
310
K_FACTOR
pulses/liter
1239
FLOWRATE
4
4.07
GPM
4.055
FLOW_VOLUME
cm3/sec
255
MASS_FLOWRATE
250
275
gm_sec
265
SEEBECK_CONSTANT
volts/C
.08
DELTA_T
0
10
°C
9.05
LOAD_EFFICIENCY_FACTOR
--
1
FRICTIONAL_WATTS
0
25
watts
20
RF_PRESSURE_DROP
psi
7.5
TOTAL_PRESSURE_DROP
psi
26.04
INPUT_POWER
-5
10000
watts
8900-10500
RF_LOAD_DC_RESISTANCE
45
55
ohms
49.6
SERIAL_NUMBER_OF_RF_LOAD
--
REFLECTION_COEFFICIENT
0
1
--
0.01 Frequency Dependent
TRANSMISSION_LOSS_FACTOR
--
0.999
CAL_FACTOR
-5
10000
watts
10000
RF_TRANS_LOSS_FACTOR
--
0.999
RF_EFFICIENCY_FACTOR
--
0.998
USER_DEF_FREQ
0
500
MHx
0 to 500 MHz
RFCALPOW
-5
10000
watts
10000
POWER_ADJ
-4
4
watts
1.25
T_PIVOT
23
27
°C
25
T_POWER
-2
2
watts/°C
-1
USE_PUMP
--
True
PUMP_SET_FREQ
44.5
45
Hz
44.8
PUMP_CONTROL
--
True
PUMP_FREQ
44.5
45
Hz
44.8
PUMP_RPM
1335
1350
rpm
1344
FLOWRATE_RPM
3.95
4.15
gpm
4.055
FLOWRATE_METER
3.9
4.1
gpm
4