1316-900 Revision AA
10kW Precision Calorimeter
4
Calorimeter/Chiller
56” height, 23” width, 30”/50”
depth (together in single bay rack (/w/table))
Calorimeter
9
” height,
19
” width,
30
” depth
(removed from rack)
Fine Chiller
11
” height,
19
” width,
25
” depth
(removed from rack)
Course Chiller (Optional)
63” height, 33” width, 30” depth
6
AC Source (Optional)
58” height, 25” width, 34” depth
6
AC Standard (Optional)
8” height, 12” width, 23” depth
6
Coolant Flow Rate
4 gallons (15.2 Liters) per minute nominal
Drift
< 50% of uncertainty over 48 hours
Input Impedance
50 Ω nominal
Communication
Ethernet
Cooling Fluid
25% DOW-Therm SR-1, 75% distilled water
Operating Temperature
68 to 86 °F (20 to 30 °C)
Storage Temperature
14 to +122 °F (-10 to +50 °C)
4
Warranty
1-year Parts & Workmanship for AE-TEGAM-manufactured components
S11 Linear Magnitude (Typical) Typical < 0.025 at
f
< 100 MHz, and <0.1 at
f
>= 100 MHz
Typical S
11
Linear Mag Performance
Optional System Components (contact your AE-TEGAM representative for more information).
AC Source
Required for routine calibration of the Calorimeter.
AC Power Meter
Required for routine calibration of the calorimeter.
Course Chiller-Air/Water Cooled Required to pre-cool cooling fluid before fine cool chiller.
P/C
For control of calorimeter and ancillary equipment.
Other
Contact TEGAM for further details and options.
1
Average of ten (10) consecutive points taken 25 seconds apart, where the standard deviation of
those points is < 0.05 W.
2
The 1316 is intended for use as a primary standard. Measurement repeatability and uncertainties
will vary among laboratories depending upon various factors, including ambient environment,
user experience, and elapsed time since the most recent system calibration. TEGAM will provide