Table B‑2. Temperature measurement specifications (Continued)
Description
Specification
Output impedance
18 Ω (for thermocouple or voltage output)
Analog output
• Analog output converter boxes (selectable): Current (0 mA
to 20 mA or 4 mA to 20 mA, linear), voltage (0 V to 5 V),
or thermocouple (type J or type K)
• Single analog converter box: (0 mA to 20 mA or 4 mA to
20 mA, linear)
• Multi-analog coverter box: Current (0 mA to 20 mA or
4 mA to 20 mA, linear), voltage (0 V to 5 V)
Additional output
10 mV/°C or 10 mV/°F to indicate temperature of sensor head.
For example, 250 mV means that the sensor head is 25°C or
25°F.
Digital output
The converter box will be equipped with one of the following
outputs:
• RS-232 or RS-485 (addressable, half-duplex), 4800 baud
to 115200 baud
• PROFINET I/O, Protocol TCP/IP, 100 Mbaud
Resolution
0.1°C (0.1°F) on the digital output
< 0.1% of the adjusted subrange on the analog output
Emissivity (Ɛ)
10% to 120%, adjustable in 0.1% increments
Response time (t
90
)
IN 600 and IN 600/5:
180 ms, switchable (0.5 s, 1 s, 2 s, 5 s, 10 s, 30 s)
IS 600, IGA 600, and IGA 600/23:
10 ms, switchable (0.5 s, 1 s, 2 s, 5 s, 10 s, 30 s)
Maximum value storage (t
cl
)
Clearing time is adjustable to: off, 0.1 s, 0.25 s, 0.5 s, 1 s, 5 s,
25 s, external, auto
Measurement uncertainty
Dependent on object
temperature T and ambient
temperature T
amb
(Ɛ = 1, t
90
= 1 s)
IN 600:
T = 0°C to 700°C: 0.7% oR or 1°C
T = 0°C to −20°C: 2°C
T = −20°C to −40°C: 3°C
IS 600, IGA 600, and IGA 600/23:
0.5% oR + 2°C
IN 600/5:
0.5% oR or 0.8°C
Advanced Energy
®
Impac
®
Series 600 Pyrometer
57010228-00C
Unit Specifications
B‑3