Table E‑1. Converter box parameters (Continued)
Parameter
Description
RS/tw
Wait time for RS-485
T/REP ON FAIL
Defines how often a request to a sensor head is repeated (in case
of no response or a wrong answer) before an error message is
shown (timeout)
Table E‑2. Sensor head parameters
Parameter
Description
S/ADDR
Sensor head address (0 to 8, 0 = no address assigned)
S/RNG BEG
Sets the beginning of the temperature subrange
☞
Important
The minimum span is 51°C (92°F).
S/RNG END
Sets the end of the temperature subrange
☞
Important
The minimum span is 51°C (92°F).
S/EMI
Emissivity (10% to 120%)
S/t90
Exposure time
IN 600 and IN 600/5:
180 ms , 0.5 s, 1 s, 2 s, 5 s, 10 s, or 30 s
IS 600, IGA 600, and IGA 600/23:
10 ms , 0.5 s, 1 s, 2 s, 5 s, 10 s, or 30 s
S/tCL
Clear time of the maximum value storage (OFF, AUTO,
EXTERN, 25 s, 5 s, 1 s, 0.25 s, 0.05 s, or 0.01 s)
Advanced Energy
®
Impac
®
Series 600 Pyrometer
57010228-00C
Pyrometer Parameters (Converter Box with Display)
E‑2