Table F‑2. Converter box and analog box UPP commands (Continued)
Parameter
Command
Description
Change IP
address of
PROFINET
module
<AA>ip<VAL>
8-digit hexadecimal ASCII. Use the
re
command to restart the pyrometer and activate
the new address.
For example, to change the PROFINET IP
address to 192.168.0.10, convert the values
to hexadecimal (C0.A8.01.0A) and send the
command,
<AA>ip<C0A8010A>
.
Read IP range
<AA>ip
?
The converter box responds with the 16-digit
hexadecimal address range.
XXXXXXXXYYYYYYYY
XXXXXXXX
: Start
YYYYYYYY
: End
Read streaming
mode
<AA>sm
The converter box responds with the streaming
mode (1-digit, ASCII decimal).
0 = Streaming off
1 = Streaming on, USB active
2 = Streaming on, RS interface active
3 = Streaming on, USB and RS interfaces
active
Change
streaming mode
<AA>sm<VAL>
Sets a new streaming mode (1 digit, ASCII
decimal). Valid values are: 0, 1, 2, or 3.
0 = Streaming off
1 = Streaming on, USB active
2 = Streaming on, RS interface active
3 = Streaming on, USB and RS interfaces
active
Read streaming
range
<AA>sm
?
The converter box responds with the streaming
range (2 digit, ASCII decimal).
01 = Standard
03 = For versions with an external RS
module
Advanced Energy
®
Impac
®
Series 600 Pyrometer
57010228-00C
UPP Commands
F‑10