3. Discover (identify) the newly connected sensor head in one of the following
ways.
◦ If using a converter box with a display, select
DIS
.
◦ In the Infra600 software, click
Discover
.
◦ Issue the 1-wire search UPP command as presented in
the Universal Pyrometer Protocols (UPP) Commands”
on page 3‑13.
Once discovered, the sensor head will then be ready for use.
☞
Important
When using an analog box, the newly connected sensor head is
immediately ready for use.
Related Links
• “Unit Connectors” on page 2‑3
Connecting Input Power to the Unit
Power your Impac pyrometer either by connecting it to a PC via USB or by
connecting it to a separate power supply. AE offers a variety of power supplies
that are compatible with your Impac pyrometer.
☞
Important
Use only one power connection (micro USB or separate power supply) when
operating the Impac pyrometer or it might lead to communication problems.
For the analog box and multi-analog box, the micro USB connection can only
be used for parameterization. For temperature measurement, use a separate power
supply connection.
For the converter box, either power connection can be used for both parameterization
or temperature measurement.
The multi-sensor box does not require separate input power.