Pyrometer Parameters
(Converter Box with Display)
Table E‑1. Converter box parameters
Parameter
Description
M/ADR
Address (device address, 00 to 97)
ADDR
Automatic addressing
• Reset all: Each sensor head will be set back to factory setting
(address 00)
• Address all: Each sensor head will be assigned a new
address (automatic addressing)
ADDR/AUTO
Activation or deactivation of automatic addressing (ON/OFF)
M/UNIT
Temperature measuring unit (Celsius or Fahrenheit)
M/DISPL
Orientation of the display
OUT/MOD
Analog output (4 mA to 20 mA, 0 mA to 20 mA, thermocouple
type J, thermocouple type K, or 0 V to 5 V
OUT/REF
Select which sensor head will be relevant for the output signal
(indicated on the display with
O
on a green background)
OUT/ERR
Show or hide errors on the display (in case of an error or wrong
value at analog out)
SP1/MOD
Select when contact shall switch (
OFF
,
BELOW
, or
OVER
) the
selected switching temperature (the temperature display turns red
when the contact has switched and is active)
SP1/REF
Select which sensor head will be relevant for the switching
contact (indicated on the display with
S
on a red background)
SP1/ERR
Show or hide errors on the display (in case of an error or wrong
value at switching contact)
SP1/TEMP
Switching temperature
SP1/HYST
Hysteresis for switching point (2°C to 20°C)
SP1/SRC
Select the source temperature for switching point
•
TEMP
= Measuring temperature of sensor head
•
TINT
= Internal temperature of sensor head
RS/INTERF
Switch between RS-232 and RS-485
RS/BAUD
Baud rate: 4800 baud to 115200 baud