E
2
T QUASAR Flare Monitors Operation Manual
Appendix A: Specifications and Parameter
Settings
•
55
Appendix A: Specifications and Parameter
Settings
Temperature Range
Designed for Flares/Pilots of any intensity
Working Distance
0 … 1320 ft. (0 … 400m)
Response Time
100 mSeconds
Outputs
Red and Green status lights, mA and set point relay contacts
Relay Contacts
120 V AC 0.5A , 24 V DC 1A , Resistive
Sensitivity
Adjustment
0 … 1 V DC
Focusing Range
20 in. (51cm) to infinity
Sighting
Visual aiming
Target Size
(Field of View)
S (dia. of circle) = D (Distance)/37.5, 60, 75, 150, or 300
Power Ratings
115
∽
± 5%, 50 / 60 Hz (standard)
220
∽
± 5%, 50 / 60 Hz (optional)
24 V
+ / - 20%, (optional)
115 Watts
Fusing
ELECTRONICS:
115
∽
.250A,’T’ Time-Lag 5x20 Little fuse type 218.250
220
∽
.125A,’T’ Time-Lag 5x20 Little fuse type 218.125
24 V
1A,’T’ Time-Lag 5x20 Little fuse type 218.001
6 882 480 Fuse, 1 A (24 V DC), (P/N 15825-1000)
6 882 490 Fuse, 4 A (24 V DC), (P/N 19154-4)
6 882 500 Fuse, 1 A (115 V AC), (P/N 19154-1)
6 882 510 Fuse, 1/4 A (115 V AC), (P/N 19534-0250)
6 882 520 Fuse, 0.5 A (230 V AC), (P/N 19154-0.5)
6 882 530 Fuse, 125 mA (230 V AC), (P/N 19534-0125)
HEATER :
115
∽
1A, Fast TR5, Wickman 370 … 1100
220
∽
.
5A, Fast TR5, Wickman 370 … 0500
24 V
4A, Fast TR5, Wickman 370 …1400
Ambient Temperature
Limits
A. -40 to 140 °F (-40 to 60 °C) with no cooling and using internal
heater.
B. If the equipment is used at the ambient temperature above
+60 °C, before turning on the equipment, the cooling system
shall be verified that the ambient temperature does not rise
above +60 °C.