Electrical Connections
All electrical connections are located in a small junction box located on the side of the larger enclosure.
Power at the specified rating must be supplied via a two wire shielded cable. A safety fuse is also
mounted in the junction box.
Do not supply power beyond the rating. Failure to do so may permanently damage the
analyzer.
The integral 4-20mA converter is internally powered and does not require external power.
DO
NOT supply any voltage to either of the two terminals of the 4-20mA output connector. Failure
to do so will permanently damage the 4-20mA converter.
Procedure
•
Remove the front cover of the junction box located on left side of the transmitters by removing
the four (4) screws securing the cover and set them aside for reinstallation.
•
Loosen the nut of the cable gland (customer provide and mounted at the bottom of the smaller
enclosure).
Two user selectable Alarms
12-24 VDC Power In
Power Fuse
4-20 mA Signal
Power Fail Alarm