The display will shift to the next higher range when the oxygen reading (actually the sensor’s signal
output) exceeds 99.9% of the upper limit of the current range. The display will shift to the next lower
range when the oxygen reading drops to 85% of the upper limit of the next lower range.
For example, if the transmitter is reading 1% on the 0-10% range and an upset occurs, the display will
shift to the 0-25% range when the oxygen reading exceeds 9.9%. Conversely, once the upset condition
is corrected, the display will shift back to the 0-10% range when the oxygen reading drops to 8.5%.
Procedure – Manual Sampling
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight MANUAL SAMPLE.
Press the ENTER key to select the highlighted menu option.
The following displays appears:
Advance the reverse shade cursor using the ARROW keys to highlight the desired RANGE.
Press the ENTER key to select the highlighted menu option.
The following display appears with the range selected and oxygen concentration of the sample gas:
3.30 %
Auto Range
0-10%
77 F
100 KPa
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
77 F
100 KPa
MANUAL RANGE
25%
10%
5%
1%
77 F
100 KPa
3.30%
Manual Range
0-10%
77 F
100 KPa
21