Select ZERO CALIBRATE. The following display appears
Allow the zero gas to flow through the transmitter until a stable oxygen reading is obtained. This may
take from several minutes to a few hours. It is highly recommended to monitor oxygen trend on a chart
recorder to ensure that the oxygen reading has stabilized; evidenced by a horizontal trend.
Press the ENTER key to ZERO calibrate or MENU key to abort and return to SAMPLING mode.
Pressing ENTER, the analyzer will complete ZERO calibration and return to Sample mode. If you press
ENTER when the oxygen reading is outside of the acceptable zero offset limit, the analyzer will fail ZERO
calibration and return to the Sample mode.
Note: The maximum zero calibration adjustment permitted is approximately 60% of the lowest full scale
range available. Thus the maximum zero calibration adjustment or zero offset for this analyzer is 0.6%
oxygen.
Factory Default Zero
The software will eliminate any previous zero calibration adjustment and display the actual signal output
of the sensor at a specified oxygen concentration. For example, assuming a zero gas is introduced, the
display will reflect an oxygen reading representing basically zero oxygen contents. However, a small
oxygen reading contributed by the sensor and electronics may be seen (typically less than 2% of the full
scale of the most sensitive range of the transmitter. This feature allows the user to test the integrity of
the sample system, sensor and electronics without removing the sensor from the sensor housing.
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATE
77 F
100 KPa
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
77 F
100 KPa
0.005%
ZERO
CALIBRTION IN PROCESS
ENTER TO CALIBRATE MENUE TO ABORT
77 F
100 KPa
PASSED
CALIBRATION
FAILED
CALIBRATION
23