R3267 Series OPT11 3GPP Level Calibration Operation Manual
5.2.1 Level Calibration
5-6
10. Repeat steps 8. through 9. for each frequency listed on Table 5-2.
Part 2 Performing Gain Cal.
11. Connect CAL. OUT and INPUT by using CBL_2 and Adap_3.
12. On the R3267 Series, press keys as follows to preset.
SHIFT, CONFIG (PRESET)
13. On the R3267 Series, press keys as follows to perform a gain calibration.
TRANSIENT, STD, Gain Cal
14. After the gain calibration sequence has completed, disconnect CBL_2 and
Adap_2.
Part 3 Performing level calibration test
15. Connect SG output and R3267 Series INPUT by using CBL_1 and Adap_3, as
shown dashed line in Figure 1.
[INPUT ATT 10 dB Setting]
16. On the R3267 Series, press keys as follows to set an input attenuator to 10 dB.
ATT, ATT AUTO/MNL(MNL), 1, 0, GHz (dB)
17. On the SG, set control as follow;
Output Frequency: 1.8483 GHz
18. On the R3267 Series, set control as follow;
Center Frequency:1.8483 GHz
19. On the R3267 Series, press keys as follows to perform auto level set function.
TRANSIENT, Modulation, Power, Tx Power, Auto Level Set
20. On the R3267 Series, after the auto level function has completed, press SINGLE
for a single Tx Power measurement.
Table 5-2 List for SG and PM Correction Data Setting
SG Output Frequency (GHz)
PM Correction Data (GHz)
1.9473
1.9473
2.0001
2.0001
2.0529
2.0529
2.0991
2.0991
2.1519
2.1519
2.1717
2.1717