92
Save Settings As… – Permits saving the chart settings to a file for later use.
Loading Settings… – Opens a file browser sub-window to select a previously
saved chart setting.
Chart Axes – Opens a Chart size control sub-window to select/deselect
Autoscale and manually rescale the chart. See figure 5-8.
Left Axis – Impedance Scale in Ohms
Bottom Axis – Range Scale in feet or meters (determined by uploaded
trace)
AutoScale ON/OFF – When ON (check in box) The software will control
the scale based on the traces data points. When OFF, user can
specify the Minimum and Maximum Scale by entering values in
those boxes for both the Ohms and Distance
To rescale the chart:
1. Turn OFF (uncheck) Autoscale for either the “Left Axis” or “Bottom Axis”
or both.
2. Enter the Minimum and Maximum Values for the desired scales.
3. Press “Apply” to use the new scales and return to the graphical display.
NOTE: To maintain the manually set scales and apply these settings to other
traces, open “Chart Window” and use the “Save Settings As” feature.
Figure 5-8
Chart Title – Rename the uploaded title and/or change title’s color
NOTE: The title only permits characters legal to use by Windows in a saved trace
name as this name will be the default name for archiving the trace. If an
uploaded plot has invalid Windows naming characters they should be replaced
before attempting to save the plot.
Chart Plot – The trace on the graph can show just the colored line (plot Color)
connecting all the data points or the colored line and the data points as small
black squares along the trace. “Tool Tip” can be turned on and set up for the
mouse pointer to show plot point data (X=Distance, Y = Impedance, or both X
and Y) when the mouse pointer is positioned on a plot point. Additionally, any
impedance trace can also be viewed in “Micro Fault” modes. See figure 5-9.