3.2.2. Custom Compensation Settings
Influencing parameters should be adjusted from their defaults when:
Environmental conditions are significantly different from default settings for emissivity, relative humidity,
temperature, and distance
High measurement accuracy is required
To change these settings, select Influence in the Target menu and then press
F2
. Ensure the setting is User; if not use the up or down button to select this
option. This enables you to change the following parameters:
Emissivity
: Choose from a list of materials or enter a value. You can
modify three “user” values by pressing the right button to enter the
value. Then press the up/down buttons to change it, and
F2
to save
the change. After you save the emissivity value, it appears in the status
bar. Allowable emissivity values are from 1.00 to 0.05.
Distance
can be set from 1.6 to 3280ft (0.5 to 999.9m). Default is 6.6ft
(2m).
Humidity
can be set from 0 to 100%. Default is 55%.
Temperature
can be set from 5° to +132°F (-15° to +55° C). Default is
55.5°F (13°C).
Note that you can also use measurement data provided by compatible Metrix ASYC IV Multimeters to
compensate for environmental factors (see §6.3.5). This requires connecting the meter to the camera via
Bluetooth (see
3.3. Tripod
To improve the sharpness and accuracy of the measurements, it is possible to use a tripod for image stabilization.
The camera can be mounted on a tripod equipped with a standard ¼-inch mounting head. Attach the camera to
the tripod by screwing the tripod mounting screw into the threaded receptacle in the battery compartment cover
on the base of the camera handle (see below).
18