F1
,
F2
, and
F3
in this screen function as follows:
F1
Search for Bluetooth devices. Devices already connected remain on the list. Those having the highest
signal levels are added and connected until the three connected devices maximum is reached.
F2
Reconnect all devices in the list.
F3
Close the Peripheral screen.
If the screen is frozen when you connect a device, measurements from the device will not appear in the
Measurements screen until you unfreeze the image and start the next image acquisition.
6.3.2. Replacing Devices
1. Disconnect the device to be removed by turning it OFF.
2. Wait until the disconnection takes effect on the camera (a message appears informing you of this).
3. Turn ON the new device to be connected.
4. With the Peripherals screen displayed, start a new search by pressing
F1
. This overwrites the list
(currently connected devices remain). The new device should be included in this list.
6.3.3. Device Polling Period
The polling period defines the frequency with which the camera “polls” the device for measurement data. To
change this:
1. In the Peripherals screen, select the device in the list and press the right button. The polling period is
highlighted in blue, with the symbol
indicating that it can be changed.
2. Use the up and down buttons to select a new polling period. Options range from 0.3s up to 50 minutes.
3. When the desired polling period is highlighted, press
F2
to save.
6.3.4. Device Measurements Display
You can view measurement data from connected devices. To do this, open the Bluetooth menu and select
Measurements
. This displays the Measurements menu.
This menu displays measurements received from the connected devices during the last acquisition.