Simple Logger
®
II (Bluetooth Edition) Series
21
3.10 Reset Switch Operation
CAUTION:
Risk of electric shock. Disconnect the unit from any input
source before opening the battery cover.
Resetting the instrument takes about 10 seconds, at which time the instru-
ment will be non responsive.
The
RESET
button resets the CPU and is located under the battery cover.
To access it, remove the battery compartment cover. The reset button is
located on the exposed PCB (see § 2.2 for location).
It is recommended to only press the
RESET
switch when the logger stops
responding to a normal press button control when not connected to Data-
View
®
. It is not recommended to reset the instrument when the logger is
recording, downloading or being configured.
If the logger is not responding to a button press, make sure the unit has
sufficient
battery power. If the battery voltage is below 1.7V, the unit will
not respond to a button press. In this condition, pressing the
RESET
button will not restart the unit.
It is recommended to download any desired session and erase the memory
before starting a new recording.
The resumption of the logger operation in the above situations assumes
that the
RESET
button cleared the fault(s). The logger will not resume
normal operation if the fault condition still exists. The instrument will try
to recover normally. However, under certain conditions, the clock and
memory full state may reset, meaning that the memory is marked as full
and the clock date/time is lost.