30
Simple Logger
®
II (Bluetooth Edition) Series
NOTE:
The native measurement of the Model L564 is voltage. This voltage is scaled to
displace current by a factor that corresponds to the users probe selection. This accuracy
specification does not account for probe errors.
MODEL
L564
ELECTRICAL
Channels
Four (two voltage and two current)
Input
Two
Three Recessed Banana Jacks
Range
0 to 1V
AC
0 to 600V
AC
Resolution
0.1mV (1mV/A scale)
0.1V
Accuracy
(50/60Hz)
0 to 10mV unspecified
10 to 50mV: ±(0.5% of R + 1mV)
50 to 1000mV: ±(0.5% of R + 0.5mV)
0 to 5V unspecified
5 to 50V: ±(0.5% of R + 1V)
50 to 600V: ±(0.5% of R + 0.5V)
Maximum Input Voltage***
5Vrms (-7.07Vpeak to +7.07Vpeak)
1.2 x 600V
Input Impedance
100K
W
16M
W
Sample Rate
64 samples/cycle
Storage Rate
Programmable from 125ms to 1 per day
Storage Modes
Start/Stop, FIFO and Extended Recording Mode
*
(XRM
™
), Store on Alarm
Recording Length
15 minutes to 8 weeks, programmable using DataView
®
Memory
1,000,000 measurements (2MB).
Recorded data is stored in non-volatile memory and retained
even if battery is low or removed.
Communication
Bluetooth (Class 2)
Power Source**
4x1.5V C-cell alkaline batteries
Battery Life
>200hrs at 125ms up to 180 days
(dependent on storage rate/recording length)
MECHANICAL
Dimensions
5.904 x 5.904 x 3.568" (150 x 150 x 91mm) w/o Sensor
Weight (with battery)
2.1lbs (0.95kg)
Cover/Case
Cover: UL94-V0 / Case: UL94-5VB
Vibration
IEC 68-2-6 (1.5mm, 10 to 55Hz)
Shock
IEC 68-2-27 (30G)
Drop
IEC 68-2-32 (1m)
ENVIRONMENTAL
Operating Temperature
14° to 122°F (-10° to 50°C)
Storage Temperature
-4° to 140°F (-20° to 60°C)
Relative Humidity
Up to 85% at 95°F (35°C), Non-condensing
Altitude
2000m
SAFETY & ELECTRO-MAGNETIC COMPATIBILITY
Safety Rating
EN61010-1; 600V CAT IV, 1000V CAT III; Pollution Degree 2
Protection Degree
IP50 (cover closed)
Electro-Magnetic
Compatibility
EN 61326-1; 07/1997 (+A1 10/1998, +A2 09/2001, +A3 05/2004)
CE Approved
Pending