32
Simple Logger
®
II (Bluetooth Edition) Series
NOTE:
When installing probes, face the arrow on the probe in the direc-
tion of the load.
Model SR193
Nominal Range
1000A
AC
for f
≤
1kHz
Measurement Range
1A to 1000A
AC
max (I >1000A not continuously)
Probe Output Signal
1mV
AC
/A
AC
Maximum Clamping Diameter
2
"
(52mm)
Safety
EN 61010-2-032, Pollution Degree 2, 600V CAT IV, 1000V CAT III
Model MN93
Nominal Range
200A
AC
for f
≤
1kHz
Measurement Range
2A to 200A
AC
max (I >200A not permanent)
Probe Output Signal
5mV
AC
/A
AC
Maximum Clamping Diameter
0.8
"
(20mm)
Safety
EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III
NOTE:
Currents <0.5A will be displayed as zero. Neutral current measures down to 0A.
Model MN193
Nominal Range
5A and 100A
AC
Measurement Range
5A: 0.005A to 6A
AC
max (1A to 1000A with ratio 1000/5 selected)
100A: 0.1A to 100A
AC
max
Probe Output Signal
5A: 200mV/A
AC
; 100A: 10mV/A
AC
Maximum Clamping Diameter
0.8
"
(20mm)
Safety
EN 61010-2-032, Pollution Degree 2, 300V CAT IV, 600V CAT III
The 5A range of the MN193 is designed to work with secondary current transformers.
Best accuracy is available when entering the transformer ratio (e.g. 1000/5A). When used to
measure 5A direct, the resolution will be limited to 0.1A max.
Specifications are subject to change without notice.